Per clinical trial (b)(4): on (b)(6) 2018, patient underwent bilateral revision augmentation mastopexy using a vertical incision.2 pieces of galashape (galaflex 3d) were used.No concomitant procedures were performed.On (b)(6) 2018, wound dehiscence, right breast, was reported.Patient was treated with medication (silvadene cream).Resolved without sequelae on (b)(6) 2018.Review by medical affairs identifies as likely related to the device and serious.
|
|
Based on the information provided, no conclusions can be made.As reported, patient experienced the post-op complication of wound dehiscence.Medical affairs has assessed the patient¿s postoperative course as being serious and likely related to the study device.Review of the device history records confirmed the device was manufactured to specification.Wound dehiscence is known inherent risk of the surgery/use of the device and included as possible complication in the adverse reaction section of the instructions for use (ifu) supplied with the device.Note: this mdr report is related to a retrospective study conducted to proactively gather real-world evidence on the use and safety of the p4hb product family, including galaflex scaffold, galaflex 3d, and galaflex 3dr.All safety events from the study are adjudicated by the bd medical monitor for an independent assessment of severity and relatedness.In an abundance of caution, events for which relatedness cannot be ruled out are processed for mdr reporting.Not returned - remains implanted.
|