• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. FRED X27; FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. FRED X27; FLOW DIVERTER Back to Search Results
Model Number FREDX4023-PMA
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stroke/CVA (1770)
Date of Event 05/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that post placement of a stent in the clinoid segment, the patient developed a right ica stroke.The proximal end of the stent was post processed with 14wire and microcatheter.The device appeared open and opposed to the vessel wall.The patient woke up and then progressively developed right ica stroke syndrome.Cta/p was negative, mri was negative, and angio was negative.Eventually the patient came back to baseline.
 
Additional Manufacturer Narrative
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device is implanted and was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.The instructions for use (ifu) identifies stroke or tia (transient ischemic attack as a potential complication associated with use of the device.
 
Additional Manufacturer Narrative
Additional information was received via email stating that the patient is "fine" and that no intervention was undertaken to address the stroke.Normal dapt protocol had been followed pre and post procedure.The vessel size was 3.27 distal, 3.78 proximal.The stent diameter versus the nominal vessel diameter was reported as 4.0 x 23 / 17 used.Angio/images and procedure notes are not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRED X27
Common Device Name
FLOW DIVERTER
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key14493365
Report Number2032493-2022-00221
Device Sequence Number6150806
Product Code OUT
UDI-Device Identifier00842429114360
UDI-Public(01)00842429114360(11)211129(17)241031(10)0000127745
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P180027
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberFREDX4023-PMA
Device Catalogue NumberFREDX4023
Device Lot Number0000127745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEADWAY 27.
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-