• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Inaccurate Delivery (2339); Mechanical Jam (2983); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a preloaded intraocular lens (iol) was scratched on the injector and did not completely exit the injector.The cartridge tip contacted the eye.Patient is in good health, but customer had to replace it with another iol.There was no vitrectomy, incision enlargement, or sutures required.No patient injury reported.Daily activities were not significantly affected.The patient status is unknown.No further information reported.
 
Additional Manufacturer Narrative
Age, weight and ethnicity: unknown/ not provided.If implanted, give date: not applicable, as lens was not implanted.If explanted, give date: not applicable, as lens was not implanted.Telephone number: (b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Additional Manufacturer Narrative
Corrected data: section h6: as part of an internal review of our mdrs it was identified that the "medical device problem code 2339 - inaccurate delivery" provided on the initial report needs to be corrected as it was incorrectly selected.The correct code that should have been used is "2983 - mechanical jam.Therefore, this supplemental filing is to correct medical device problem code 2339 as it is not applicable.The following field was updated accordingly: section h6: medical device problem code: 2983 - mechanical jam.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section d9.Device available for evaluation? yes.Returned to manufacturer on dec 21, 2022.Section h3.Device evaluated by manufacturer? yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the complaint handpiece was received with the lens stuck inside of the cartridge and overridden by the plunger rod.The lens could be observed to be damaged inside of the cartridge.The handpiece assembly was inspected and, no issues could be observed that could cause or contribute the complaint issue.The lens was removed from the cartridge and the lens could be observed to be damaged and in several pieces.The observed issue 'lens damaged' is similar to the complaint issue 'cosmetic issues' and could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.The complaint issue of 'stuck in cartridge' was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issue during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14493425
Report Number3012236936-2022-01352
Device Sequence Number5837332
Product Code HQL
UDI-Device Identifier05050474558328
UDI-Public(01)05050474558328(17)240105
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2024
Device Model NumberPCB00
Device Catalogue NumberPCB0000230
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer10/20/2022
01/03/2023
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date10/20/2022
01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-