• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40E
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age, weight and ethnicity: per eu regulation 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted; give date: the intraocular lens was inserted and removed during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was inserted and removed during the same procedure.Telephone number: (b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was defective and that the surgeon explanted it because the posterior chamber broke.Through follow-up, it was learnt that there was a rupture of the posterior chamber and the lens was removed during the initial procedure, because there was lack of support.Neither the lens nor any other johnson & johnson device is reported to have caused or contributed to the capsule tear and the surgery could not be completed with a back-up lens.However, this could not be verified.No further information was provided.
 
Additional Manufacturer Narrative
Additional information: section d9 - device available for evaluation? yes section d9 - date returned to manufacturer: 13th may 2022 section h3 - device evaluated by manufacturer? yes device evaluation: the complaint lens was received inside of the original lens case.Visual inspection under magnification revealed that the lens was received with two cuts into the optic body.The lens was cleaned and both haptics were observed to be bent.However, no issues that could cause or contribute to a capsular tear could be identified.The complaint issue dc-lens damaged was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14493535
Report Number3012236936-2022-01326
Device Sequence Number14921426
Product Code HQL
UDI-Device Identifier05050474502161
UDI-Public(01)05050474502161(17)240620
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2024
Device Model NumberAR40E
Device Catalogue NumberAR40E00230
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-