• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENERGEN CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENERGEN CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number N140
Device Problems Premature Discharge of Battery (1057); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2022
Event Type  Injury  
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Further, the device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned and analysis will be performed, a supplemental report will be filed if there is any further relevant information from that review.
 
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned and analysis will be performed, a supplemental report will be filed if there is any further relevant information from that review.The returned device was thoroughly inspected and analyzed.Review of device memory confirmed that a low voltage alert (code 1003) was recorded.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised low voltage capacitors.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Investigation has determined that the low voltage capacitors can become compromised due to the presence of excess hydrogen gas within the device case.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Further, the device was explanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENERGEN CRT-D
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14493705
MDR Text Key292566650
Report Number2124215-2022-14095
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526480959
UDI-Public00802526480959
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S255
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/24/2016
Device Model NumberN140
Device Catalogue NumberN140
Device Lot Number116140
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age84 YR
Patient SexMale
Patient Weight73 KG
-
-