STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 3.5 ACCOLADE (127 DEG) SIZE 3.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-3535
Medical Device Problem Codes Degraded (1153); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Metal Related Pathology (4530)
Date of Event 03/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2012 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2012 and was revised on (b)(6) 2022.It is further alleged that he suffered injuries as a result of implantation and explantation of the device(s) at issue, device recall and excessive levels of chromium and cobalt in his blood.Update based on medical review dated 09 july 2022.This inquiry reports the case of femoral head disassociation and erosion of the femoral trunnion.This occurred approximately 10 years after initial implantation.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation and abnormal ion level involving an accolade stem was reported.The event was confirmed through clinician review of the provided medical records.Method & results: -device evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of the provided medical records by a clinical consultant indicated: this inquiry reports the case of femoral head disassociation and erosion of the femoral trunnion.This occurred approximately 10 years after initial implantation.The implants themselves were well fixed.The patient did have symptoms for over two years prior to seeking attention which possibly could have contributed to this failure.Review of the x-rays revealed the acetabular component to be in neutral version.In my experience this can cause impingement in flexion which also could contribute to this failure.This failure is related to the articulation between the femoral head and femoral trunnion.No other implants are involved in this phenomena.I can confirm that this event occurred since i was able to review the pre-revision xray.The root cause of femoral head disassociation and femoral trunnion erosion is multifactorial including surgical technique, patient factors such as bmi and activity level, and implant factors.-device history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusion: a review of the provided medical records by a clinical consultant indicated: this inquiry reports the case of femoral head disassociation and erosion of the femoral trunnion.This occurred approximately 10 years after initial implantation.The implants themselves were well fixed.The patient did have symptoms for over two years prior to seeking attention which possibly could have contributed to this failure.Review of the x-rays revealed the acetabular component to be in neutral version.In my experience this can cause impingement in flexion which also could contribute to this failure.This failure is related to the articulation between the femoral head and femoral trunnion.No other implants are involved in this phenomena.I can confirm that this event occurred since i was able to review the pre-revision xray.The root cause of femoral head disassociation and femoral trunnion erosion is multifactorial including surgical technique, patient factors such as bmi and activity level, and implant factors.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Product surveillance will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 3.5 ACCOLADE (127 DEG) SIZE 3.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key14493771
Report Number0002249697-2022-00754
Device Sequence Number14895215
Product Code LPH
UDI-Device Identifier04546540510754
UDI-Public04546540510754
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2017
Device Model Number6021-3535
Device Catalogue Number6021-3535
Device Lot Number39134101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2022
Supplement Date Received by Manufacturer07/25/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-