• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I.T.S. GMBH CANCELLOUS STABILIZATION SCREW, D=3.0MM, L=16MM, RH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I.T.S. GMBH CANCELLOUS STABILIZATION SCREW, D=3.0MM, L=16MM, RH Back to Search Results
Model Number 37303-16
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have inspected the qc documents of the lot and the quality inspection forms as well as the material certificates showed no deviations and complies with the specifications.The quality forms met all dimensional and visual criteria at the time of release with no issues documented during the manufacturing that would contribute to this complaint condition.A review of the raw material device history record revealed no deviation.The material was determined to be conforming and was used as is per product development approval.The angulation of the screw was aimed towards the proximal radial head and did not seem to exceed the 15 degree variable angle.But, the screw was already locked in the plate hole and the surgeon tried further and caused the head of the screw to crack.
 
Event or Problem Description
It was reported that the head of a locking screw cracked during final tightening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANCELLOUS STABILIZATION SCREW, D=3.0MM, L=16MM, RH
Common Device Name
CANCELLOUS STABILIZATION SCREW, D=3.0MM, L=16MM, RH
Manufacturer (Section D)
I.T.S. GMBH
autal 28
lassnitzh[?]he, styria 8301
AU  8301
Manufacturer (Section G)
I.T.S. GMBH
autal 28
lassnitzh[?]he, styria 8301
AU   8301
Manufacturer Contact
florian grill
autal 28
lassnitzhoehe, styria 8301
AU   8301
MDR Report Key14494157
Report Number3004369035-2022-00009
Device Sequence Number20335765
Product Code HWC
UDI-Device Identifier09120047295014
UDI-Public09120047295014
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K033756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number37303-16
Device Catalogue Number37303-16
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-