Model Number 6191-1-001
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 08/24/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Patient had a revision for aseptic loosening of a s&n baseplate cemented with stryker cement.
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving simplex cement mix was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no relevant records pertaining to this event were provided for review.Product history review of the batch manufacturing records indicate packs were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar event for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Patient had a revision for aseptic loosening of a s&n baseplate cemented with stryker cement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLEX P FULL DOSE 1 PACK
Common Device Name
BONE CEMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
sanjana talathi
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14494457
Report Number0002249697-2022-00757
Device Sequence Number12161173
Product Code LOD
UDI-Device Identifier07613327128444
UDI-Public07613327128444
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K062553
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model Number6191-1-001
Device Catalogue Number6191-1-001
Device Lot NumberRJX135
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer12/12/2023
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date01/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/26/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age53 YR
Patient SexMale
Patient Weight112 KG
-
-