As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 07/2024).
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Two photos were reviewed.The first photo shows the catheter having a bond joint break between the balloon and catheter.No other anomalies noted.The second photo shows incident report of the complaint.Therefore, based on the photos review the reported balloon rupture and difficult to remove will be remains inconclusive.However, the identified bond joint break can be confirmed.Therefore, the investigation for the reported balloon rupture and difficult to remove remains inconclusive as no evidence could noted on the provided photos.However, the investigation is confirmed for the identified break as a bond joint failure between the catheter and balloon was noted on the submitted photos.The identified bond joint break might be contributed to the reported difficult to remove.However the definitive root cause for the reported balloon rupture and difficult to remove and identified break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date: 07/2024), g3, h6 (device).H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|