• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR8054
Medical Device Problem Codes Break (1069); Difficult to Remove (1528); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 07/2024).
 
Event or Problem Description
It was reported that during an ultrasound guided angioplasty procedure via the right upper arm, the pta balloon was allegedly ruptured at the distal end.It was further reported that the device was difficult to remove from the introducer sheath.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.Two photos were reviewed.The first photo shows the catheter having a bond joint break between the balloon and catheter.No other anomalies noted.The second photo shows incident report of the complaint.Therefore, based on the photos review the reported balloon rupture and difficult to remove will be remains inconclusive.However, the identified bond joint break can be confirmed.Therefore, the investigation for the reported balloon rupture and difficult to remove remains inconclusive as no evidence could noted on the provided photos.However, the investigation is confirmed for the identified break as a bond joint failure between the catheter and balloon was noted on the submitted photos.The identified bond joint break might be contributed to the reported difficult to remove.However the definitive root cause for the reported balloon rupture and difficult to remove and identified break could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date: 07/2024), g3, h6 (device).H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during an ultrasound guided angioplasty procedure via the right upper arm, the pta balloon allegedly ruptured at the distal end.It was further reported that the device was difficult to remove from the introducer sheath.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14494705
Report Number2020394-2022-00405
Device Sequence Number17646950
Product Code LIT
UDI-Device Identifier00801741058493
UDI-Public(01)00801741058493
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR8054
Device Catalogue NumberDR8054
Device Lot Number93UF0105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer06/24/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexFemale
-
-