|
CONMED UTICA FRAZIER, WITH CNTL VENT + OBTR, 12 FR.; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
|
Back to Search Results |
|
| Model Number |
0033120 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, 0033120 , for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Received two 0033120 in original packaging.Lot number was verified.Performed a visual inspection of the device, there were no obvious signs of a breach in the seal however, the device was encroaching into the seal.Performed a functional inspection, the devices were dye leak tested per tm-10-217 rev.F which indicated that the packaging had an insufficient heat seal on one of the two packages.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 55 complaints, regarding 135 devices, for this device family and failure mode.During this same time frame 18,707,120 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.000007.Conmed encourages the inspection and/or test of all medical equipment prior to use to ensure all devices are functioning as expected.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Search Alerts/Recalls
|
|
|