| Model Number |
PC9000 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the handheld device was left on a patient in error.As a result, the light from the video processor illuminator caused a small hole to be burned into the patient drape.There was no adverse consequences to the patient as a result.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
The device manufacturer date is not known.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: drape burn probable root cause: drape caught between camera front plate and drape lens the reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the handheld device was left on a patient in error.As a result, the light from the video processor illuminator caused a small hole to be burned into the patient drape.There was no adverse consequences to the patient as a result.
|
| |
|