Model Number 670185
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Perforation (2205)
Health Effect - Clinical Code Choking (2464)
Date of Event 04/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was unable to be evaluated as it was not returned.If the device is received in the future an evaluation will be performed.The reported event was unable to be confirmed or replicated.A cause of the reported event could not be determined at this time.
 
Event or Problem Description
It was reported that the trach tube's cuff was filled with 9ml of water, and the user noticed that patient was choking.User tried to draw the 9ml from the cuff, but there was only 3ml.The cuff was pierced.Additional information received: the physician changed the trach tube to resolve the situation, in addition to cough assist being provided by physiotherapist to patient.
 
Additional Manufacturer Narrative
Other, other text: this mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, ADULT
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14496902
Report Number3012307300-2022-09617
Device Sequence Number12264303
Product Code JOH
UDI-Device Identifier15021312006087
UDI-Public15021312006087
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/14/2025
Device Model Number670185
Device Catalogue Number670185
Device Lot Number4089239
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer03/22/2023
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date04/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
Patient Weight86 KG
-
-