• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HICKMAN DUAL LUMEN CV CATHETER, 7F; CHRONIC CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HICKMAN DUAL LUMEN CV CATHETER, 7F; CHRONIC CATHETERS Back to Search Results
Model Number 0600574
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code Air Embolism (1697)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 09/2025).Device pending return.
 
Event or Problem Description
It was reported that during a chronic catheter placement procedure in the right internal jugular, the patient allegedly experienced air embolism.It was further reported that medical interventions was required to remove air embolism through medications.The patient reportedly to be stable.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.The instructions for use states that, warning: do not use the catheter if there is any evidence of mechanical damage or leaking.Accessories and used in conjunction with this device should incorporate luer lock connectors.Avoid accidental device contact with sharp instruments and mechanical damage to the catheter material.Use only smooth edge atraumatic clamps or forceps.Avoid perforating, tearing or fracturing the catheter when using a guidewire.Prior to beginning placement procedure, do the following: examine package carefully before opening to confirm its integrity and that the expiration date has not passed.The device is supplied in a double sterile package and is non-pyrogenic.Do not use if package is damaged, opened or the expiration date has passed.(product is) sterilized by ethylene oxide.Do not resterilize.Inspect kit for inclusion of all components.Do not use the catheter if there is any evidence of mechanical damage or leaking.Avoid sharp or acute angles during implantation which could compromise the patency of the catheter lumens.If sutures are used to secure the catheter, make sure they do not occlude or cut the catheter.Clamping the catheter.Selection of the catheter clamp is very important since the catheter is vital to your care.The wrong clamp can damage the catheter.Follow these three rules for clamping: 1.Use only smooth-edged clamps.2.Always clamp the catheter over the reinforced clamping sleeve or tape tab, as instructed by your nurse.Never clamp over the reinforced segment directly adjacent to the connector.(see diagram) 3.Follow the directions of your doctor or nurse regarding when to clamp.Most hickman and broviac catheters come with pre-attached clamps and reinforced clamping sleeves - do not clamp here.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during a chronic catheter placement procedure, the device allegedly had bubbles in the line.The procedure was completed using another device.Reportedly medical intervention required for air embolism.The current patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN DUAL LUMEN CV CATHETER, 7F
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14496965
Report Number3006260740-2022-01946
Device Sequence Number6150891
Product Code LJS
UDI-Device Identifier00801741051876
UDI-Public(01)00801741051876
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600574
Device Catalogue Number0600574
Device Lot NumberHUEX1783
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/20/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-