DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRIO MESH; SURGICAL MESH Back to Search Results
Catalog Number 0010212
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Bacterial Infection (1735); Necrosis (1971); Seroma (2069)
Date of Event 08/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, about 5 years post-implant of bard/davol ventrio mesh, the patient was diagnosed with seroma, mesh infection and necrosis for which the mesh was partially explanted.The outcome of this aes are reported as recovering.The clinician has assessed the patient¿s postoperative course as being "definitely" related to the study device and definitely related to the index procedure, however, based on the information provided, no conclusion can be made.Mesh infection and seroma are known inherent risk of hernia repair surgery.The instructions-for-use, supplied with the device lists infection and seroma as possible complications.In regards to the infection, the warning section of the ifu states "if an infection develops, treat the infection aggressively.Consideration should be given regarding the need to remove the mesh.An unresolved infection may require removal of the device." review of manufacturing records confirms product was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of 195 units released for distribution in july, 2015.This mdr represents the 2015 implanted ventrio mesh (device #2).Additional mdr was submitted to represent the 2013 implanted ventrio mesh (device #1).Not returned - partially explanted.
 
Event or Problem Description
Alleged per clinical trial (b)(6): on (b)(6) 2013, the patient underwent multi-recurrent open ventral incisional hernia repair, during which a bard/davol ventrio hernia patch (device #1) was placed.Fascia and full skin closure was achieved.On (b)(6) 2015, patient experienced incarcerated recurrent incisional hernia.On (b)(6) 2015 patient admitted to the hospital.The recurrence resulted in reoperation.Patient underwent open ventral incisional hernia repair, during which a bard/davol ventrio hernia patch (device 2) was placed.As reported, no action was taken related to the previously implanted mesh (device 1).The patient was discharged to home on (b)(6) 2015.The ae (recurrence ¿ device 1) is clinically assessed to be serious and possibly related to the study device and possibly related to the procedure.(b)(6) 2019, a culture was obtained the culture tested positive for bacterial infection as gordonia.On (b)(6) 2020, patient developed chronic nonhealing abdominal wall with soft tissue necrosis and seroma (ae 2).The reported ae resulted in reoperation and partially removal of both mesh.On (b)(6) 2020, patient underwent additional surgical intervention, which included additional partial explant of both mesh, related to ¿chronic mesh infection¿ (ae 2).Ae 2 is clinically assessed to be definitely related to the study devices and definitely related to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRIO MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14497373
Report Number1213643-2022-00339
Device Sequence Number5799078
Product Code FTL
UDI-Device Identifier00801741016394
UDI-Public(01)00801741016394
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K100229
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source Other,Study,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/28/2018
Device Catalogue Number0010212
Device Lot NumberHUZF0068
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/24/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-