| Model Number |
PBD-1033-0707 |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
04/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device has not been returned to olympus for evaluation.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
| |
|
Event or Problem Description
|
|
The customer reported during an ercp to remove a stent, the physician encountered resistance when pushing the grasper down the scope.The physician pushed harder and a piece of blue plastic came out of the end of the scope.The scope was removed and an alternative scope was used to complete the patient's procedure as planned.An x-ray was taking for this patient and they are unable to confirm if the end of the pigtail stent (also blue) was missing in the patient.The blue plastic fragment released from the first scope was retrieved.It is unknown how it was retrieved.The facility reviewed the use history of the scope and determined it was last used to implant a pigtail stent for a different patient.This event includes 2 reports: (b)(6): tjf-260v, scope.(b)(6): pbd-1033-0707, stent.This is report 2 of 2 for (b)(6): pbd-1033-0707, stent.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information from the customer, the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.H4: 2021july.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to a tensile load greater than the stent's proof stress was applied to the stent.However, we were unable to identify the cause of the occurrence because we were unable to check the actual product.The instruction manual contains the following descriptions.This is possible to prevent the reported event.Drawing number: gk6896 revision number:07 when retrieving this instrument from the bile duct, grasp a part avoiding the vicinity of side hole or side flap as much as possible, and slowly aligned with bile duct withdraw it with the endoscope observing fluoroscopic images.If a part around the side hole or side flap is strongly grasped by a snare or grasping forceps, or the stent is withdrawn with excessive force, this instrument may break and leave debris in the bile duct.Do not retrieve this instrument in other methods than described in this instruction manual.Otherwise, it could cause migrating or falling off from the papilla and could damage the instrument.To retrieve the stent, use grasping forceps fg-44nr-1 in accordance with its instruction manual.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
Additional information received from the customer: the customer stated the device fragment fell outside of the patient but also stated the fragment was retrieved from the patient.The customer also stated there was no patient harm, injury or infection.
|
| |
|