• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LENS INTEGRATED SYSTEM WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LENS INTEGRATED SYSTEM WIFI VERSION; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72204354
Medical Device Problem Code Optical Discoloration (2999)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use, the color image of the lens integrated system was wobbling, there was complete color change in the monitor.The procedure was completed with the same device.There was a delay greater than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H10: h3, h6: the reported device was not returned to the designated complaint facility for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Factors that could have contributed to the reported event include failure of a concomitant device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately.There were no indications to suggest the anticipated risk is not adequate.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H11: corrected data g2 (company representative should be checked).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS INTEGRATED SYSTEM WIFI VERSION
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14497769
Report Number1643264-2022-00179
Device Sequence Number12170871
Product Code GCJ
UDI-Device Identifier00885556617878
UDI-Public00885556617878
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K153606
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72204354
Device Catalogue Number72204354R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-