As reported, following a percutaneous transluminal angioplasty (pta) and using a micropuncture transitionless stiffened cannula access set, the introducer shaft tore off the head of the introducer when attempting to remove the device.The access site was the femoral artery, and it was very scarred and calcified.The initial puncture went as planned, and there were no issues while inserting the wire and sheath.The initial wire was manipulated and/or removed through the entry needle.When attempting to remove, the physician felt resistance and switched to another manufacturer's guidewire and it came loose.A section of the device did not remain inside the patient¿s body.The patient did not require additional procedures due to this occurrence.According to the initial reporter, the patient did not experience adverse effects due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, following a percutaneous transluminal angioplasty (pta) and using a micropuncture transitionless stiffened cannula access set, the introducer shaft tore off the head of the introducer when attempting to remove the device.The access site was the femoral artery, and it was very scarred and calcified.The initial puncture went as planned, and there were no issues while inserting the wire and sheath.The initial wire was manipulated and/or removed through the entry needle.When attempting to remove, the physician felt resistance and switched to another manufacturer's guidewire, and it came loose.A section of the device did not remain inside the patient¿s body.The patient did not require additional procedures due to this occurrence.According to the initial reporter, the patient did not experience adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.One used outer catheter was returned to cook for investigation, along with the separated hub.The shaft was found to be separated, leaving shaft material inside of the hub.The length of the separated section was 9.9-centimeters.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu cautions ¿do not attempt to insert or withdraw the wire guide and/ or introducer if resistance is felt.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that the patient¿s scarred and calcified anatomy contributed to this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|