As reported by the field, during a coil embolization of the carotid artery, a 3mm x 6cm deltaxsft10 coil (dlx100306, 30481268) was attempted to be placed and would only advance halfway into the echelon 10 microcatheter (mc).It was not necessary to remove or replace the mc.Neither the coil or the mc appeared to be damaged.Nothing was obstructing the microcatheter.The procedure was completed by using a competitor¿s coil.There was no patient injury reported.Based on the device analysis of the product received, the embolic coil was found slightly stretched.
|
|
Product complaint # (b)(4).The purpose of this mdr submission is to report the findings of the device investigation.The embolic coil was found stretched, meeting regulatory reporting criteria.Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Complaint conclusion: as reported by the field, during a coil embolization of the carotid artery, a 3mm x 6cm deltaxsft10 coil (dlx100306, 30481268) was attempted to be placed and would only advance halfway into the echelon 10 microcatheter (mc).It was not necessary to remove or replace the mc.Neither the coil or the mc appeared to be damaged.Nothing was obstructing the microcatheter.The procedure was completed by using a competitor¿s coil.There was no patient injury reported.A non-sterile deltaxsft 10 3mm x 6cm was received for analysis on a pouch.The device was noted to be in good normal conditions; all the components were returned with no apparent damage such as kinks, and separations, the introducer was not kinked.The coil component was attached to the rest of the unit.The echelon microcatheter was returned along with the three (3) complaint coils.The coil component was inspected under microscopic magnification, and it was found to be slightly stretched.The functional test could not be performed due to the conditions noted on the coil component.A manufacturing record evaluation (mre) was performed for the finished device [30481268] number, and no non-conformances related to the reported complaint condition were identified.The device was returned to cerenovus for further evaluation.Upon arrival, the device was noted to be in good normal conditions; all the components were returned with no apparent damage such as kinks, separations, the introducer was not kinked.The coil component was attached to the rest of the unit.The echelon microcatheter was returned along with the three (3) complaint coils.The coil component was inspected under microscopic magnification, and it was found to be slightly stretched.Although the echelon microcatheter was noted to be in good conditions, the condition noted on the coil component precluded further testing.The deltaxsft 10 3mm x 6cm systems are compatible with microcatheters with inner lumen diameters ranging from 0.0165 to 0.017 in (0.419 to 0.432 mm).The concomitant microcatheter has a distal inner diameter of 0.017 in (0.432 mm), which is within the recommended range.Coil stretching is a known potential issue associated with the use of this device.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issue from occurring.Stretching can occur during procedure handling where force may have been inadvertently applied.The complaint documented that there was resistance felt between the coil and the concomitant microcatheter; the complaint coil was removed from the patient.It is possible that undue force may have inadvertently been applied during the coil removal from the microcatheter which resulted in the coil stretching.The event described was confirmed based on the damage noted on the coil.Although the device could not be advanced through the microcatheter, such damage seems to be secondary to the impeded condition encountered during the procedure, which ultimately might have prompted the user to inadvertently apply force to remove the coil from the microcatheter.The complaint documented that the three (3) coils got stuck at the midpoint of the microcatheter, but the microcatheter was received without apparent damage; it is possible that the position of the microcatheter inside the patient, or the patient¿s anatomy might have contributed to the failure during the procedure, however, this cannot be conclusively determined.With the limited information available, a root cause cannot be assigned, the failure reported may relate to factors not described within the information available.There is no indication that the issue reported in the complaint is a result of a defect inherently related to the device.A manufacturing record evaluation was performed, and no non-conformance was found during the review.It should be noted that product issue could be caused by multiple factors, the instructions for use (ifu) does contain the following caution: if the microcoil system becomes immobile in the infusion microcatheter, apply a gentle push-pull motion to free it.If unsuccessful, remove both microcatheter and microcoil system together as a unit and replace with new devices.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|