It was reported a safe-t-j fixed core wire guide was damaged during a drain insertion procedure.When removing the metal stylet of the catheter, the wire was reportedly "severely kinked, stretched, and unable to be removed from the inside of the stylet." this made the device unusable.There was no patient harm as a result of this occurrence.
|
|
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Investigation ¿ evaluation.On (b)(6) 2022, cook medical inc.Received a complaint from yuyan li, a representative at the (b)(6) hospital, located in the city of (b)(6) reporting the following: on (b)(6) 2022, during an initial insertion of the multipurpose drainage catheter, the safe-t-j fixed core wire guide (rpn: tscf-35-80-3, lot:14382168) became stuck.The wire could not easily be removed from the metal stylet and catheter while in place.The wire appeared to have snapped and uncoiled.Upon the removal of the stylet / catheter combo, it was discovered the wire was broken, resulting in coil elongation, without separation.A review of the complaint history, device history record, manufacturing instructions, quality control procedures, and specifications of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also completed a review of the device history record (dhr).The dhr for lot# 14382168, discovered a total of 6 recorded nonconformances.A review of the recorded nonconforming criteria, determined that "bent / kinked" and "weld, large" were possibly relevant to the reported difficulty.All devices were scrapped prior to further processing.To date, a further search of our database records revealed one additional complaint having the same lot number and from the same reporting facility and was reported for "wire guide getting caught on the 6.0 fr drainage catheter".Since there is objective evidence the dhr was fully executed, with no other lot-related complaints being received from another facility, cook has concluded that there is no evidence that non-conforming product exists in house or in the field and that device was manufactured to current specification.Based on the information provided, no returned device, and the results of the investigation, it was determined the cause of this event is traced to a component failure without a manufacturing or design deficiency.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|