Model Number 25-878-05-91
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Multiple mdr reports were filed for this patient.Mdr: 9610905-2022-00023.
 
Event or Problem Description
It was reported that screws became loose due to patient grinding their teeth.They were removed and replaced.
 
Additional Manufacturer Narrative
An investigation was performed in the lab and there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.Review of the device history records was not possible due to no lot number being identified.The investigation results conclude that the root cause for this failure could not be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KLS MARTIN L.P.
p.o. box 16369
jacksonville FL 32245
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key14498335
Report Number9610905-2022-00022
Device Sequence Number14847450
Product Code HRS
UDI-Device Identifier00888118047694
UDI-Public(01)00888118047694
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K971297
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-878-05-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age26 YR
Patient SexMale
Patient Weight115 KG
-
-