• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. ALTERA ARTICULATING EXPANDABLE TLIF; ALTERA INSERTER SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. ALTERA ARTICULATING EXPANDABLE TLIF; ALTERA INSERTER SLEEVE Back to Search Results
Model Number 6124.0002
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Laceration(s) (1946)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery a vein was punctured causing the patient significant blood loss.
 
Additional Manufacturer Narrative
All four components were returned for evaluation.Initial observation shows the medial tip of the altera inserter sleeve is slightly deformed outward and some wear is visible on the distal tip.This type of damage is seen when the user excessively tightens the gold knob on the altera inserter base.Additional information provided that the case was an l5-s1 tlif with disc prep and mis screws done with robotic navigation.The altera spacer was articulated without issue.When the surgeon attempted to medialize the implant, the implant and inserter assembly shifted through the anterior longitudinal ligament (all) of the l5-s1 disc space.It was reported there was an osteophyte near the disc space on the anterior portion of the l5 vertebra that may have caused or contributed to the all rupturing, which allowed the implant to contact the vein.It was also reported that the implant was not able to be collapsed to remove.This is likely due to pulling on the inserter, as this causes the spring in the threaded shaft to compress and disengage the driver shaft.After removing the implant, it was noted that there were pieces of a mesh stent inside the altera implant, and the patient started to bleed significantly.It is possible that the osteophyte could have compromised the integrity of the all allowing the event to occur; however, an exact cause to the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTERA ARTICULATING EXPANDABLE TLIF
Common Device Name
ALTERA INSERTER SLEEVE
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key14498442
Report Number3004142400-2022-00075
Device Sequence Number14733402
Product Code LXH
UDI-Device Identifier00889095174557
UDI-Public00889095174557
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140411
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6124.0002
Device Lot NumberFMX021KC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-