MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOLDING SLEEVE; SURGICAL SCREWDRIVER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOLDING SLEEVE; SURGICAL SCREWDRIVER, REUSABLE Back to Search Results
Model Number 313.97
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the sales rep that during an unknown procedure on (b)(6) 2022, it was observed that the holding sleeve from the imf screw set would not grasp imf screw.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.This complaint involved one(1) device.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a j&j sales representative.The lot number was unknown.Therefore, the expiration date and device manufacture date were unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6 investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the holding sleeve was found the device loose.This can be related to an inner failure of the compression spring component, which agrees with the reported complaint condition.No other defect was observed.A dimensional inspection for the holding sleeve was unable to be performed due to post manufactured damage.A functional test was not conducted since the mating device was not returned as part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the complex geometry of the device would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: yes, no issues.Dimensional inspection: n/a.Device history: due to the age of more than 10 years of the complained device a wear or use related root cause is the most likely reason of the complained malfunction.Per franchise complaint product investigation procedure is for complaints for which a non-manufacturing related probable cause has been identified no manufacturing record evaluation is required.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLDING SLEEVE
Common Device Name
SURGICAL SCREWDRIVER, REUSABLE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14498519
Report Number2939274-2022-01919
Device Sequence Number5106593
Product Code HXX
UDI-Device Identifier10887587037196
UDI-Public10887587037196
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number313.97
Device Catalogue Number313.970
Device Lot NumberA4GM779
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2000
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age17 YR
Patient SexFemale
-
-