ZIMMER TMT NEXGEN COMPLETE KNEE SOLUTION, TRABECULAR METAL STD PRIMARY PATELLA; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Scar Tissue (2060)
Date of Event 10/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Patient developed arthrofibrosis after a right tka on (b)(6) 2020 and required a manipulation under anesthesia.According to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthopaedics and related research (how to treat the stiff total knee arthroplasty? a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is related to limited range of motion and stiffness due to adhesions/scar tissue.Additional associated products & mdrs (b)(4) femur cemented cruciate retaining(cr) narrow rt sz 7 lot# 64633161, mdr: 3007963827-2022-00141; (b)(4) articular surface fixed bearing(uc) right 11 mm lot# 64418973, mdr: 3007963827-2022-00143.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
Patient received an initial right total knee arthroplasty.Subsequently, underwent a manipulation, under anesthesia of the right knee two months post procedure, due to arthrofibrosis of the joint.No further complications were reported.Patient demonstrated overall improvement with pain, satisfaction, and quality of life and completed the study per protocol with no further interventions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN COMPLETE KNEE SOLUTION, TRABECULAR METAL STD PRIMARY PATELLA
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer (Section G)
ZIMMER TMT
10 pomeroy road
parsippany NJ 07054
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14498747
Report Number3005751028-2022-00014
Device Sequence Number20090823
Product Code JWH
UDI-Device Identifier00889024465664
UDI-Public(01)00889024465664(17)300630(10)64635672
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K011904
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00587806529
Device Lot Number64635672
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight74 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-