• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C2DX, INC. T/PUMP PROFESSIONAL; T/PUMP, TP700

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C2DX, INC. T/PUMP PROFESSIONAL; T/PUMP, TP700 Back to Search Results
Model Number TP700
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Partial thickness (Second Degree) Burn (2694)
Date of Event 04/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On 13-may-2022, the c2dx service technician performed functional checks according to the t/pump professional maintenance manual.The flow and temperature tests were both within specification.While running maintenance manual functional checks, the service technician noticed corrosion of the brass block and identified other preventative maintenance items to be addressed before potential return to service.
 
Event or Problem Description
(b)(6) emailed c2dx on 5/4/2022 indicating that a burn incident had occurred at a location that agiliti health services, referred to as (b)(6) hospital.The patient was undergoing a four-hour surgery around noon on (b)(6), and the t/pump was being used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T/PUMP PROFESSIONAL
Common Device Name
T/PUMP, TP700
Manufacturer (Section D)
C2DX, INC.
555 e. eliza street
ste a
schoolcraft MI 49087
Manufacturer (Section G)
C2DX, INC
555 e. eliza street
ste a
schoolcraft MI 49087
Manufacturer Contact
ryan eggleston
555 e. eliza street
ste a
schoolcraft, MI 49087
8889022239
MDR Report Key14500587
Report Number3015489752-2022-00001
Device Sequence Number14892564
Product Code ILO
UDI-Device Identifier07613327169249
UDI-Public07613327169249
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K073675
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTP700
Device Catalogue NumberTP700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-