• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RXSIGHT, INC. LIGHT ADJUSTABLE LENS (LAL) Back to Search Results
Model Number 60005
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 04/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Site reported the light adjustable lens was explanted as the patient was unhappy with the visual outcome.Another lal was implanted during the exchange on (b)(6) 2022.Rxsight's first awareness of this event was 4/27/2022.Device history record for the lens was reviewed.No issues were noted.The explanted lens was returned for evaluation.The lens was cut into multiple fragments.Visual and optical testing of the lens found no issue with the lens.
 
Event or Problem Description
Site reported the light adjustable lens was explanted as the patient was unhappy with the visual outcome.Another lal was implanted during the exchange on (b)(6) 2022.Rxsight's first awareness of this event was 4/27/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS (LAL)
Common Device Name
LIGHT ADJUSTABLE LENS (LAL)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
kelli carrillo
100 columbia
aliso viejo, CA 92656
MDR Report Key14501475
Report Number3012712027-2022-00019
Device Sequence Number5124187
Product Code PZK
UDI-Device Identifier00818806020746
UDI-Public01008188060207461121060117240531
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model Number60005
Device Catalogue Number60005
Device Lot NumberL07-003254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2022
Initial Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-