• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CVC SET: 3-LUMEN 7 FR X 20 CM; CATHETER PERCUTANEOUS Back to Search Results
Model Number IPN914770
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported the needle hub broke and separated during insertion.No patient harm.The device was replaced.The patient's condition is reported as fine.
 
Event or Problem Description
It was reported the needle hub broke and separated during insertion.No patient harm.The device was replaced.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one introducer needle and lidstock for analysis.Signs of use in the form of biological material were present on the needle cannula.Visual analysis revealed that the needle hub was broken and separated.A portion of the distal end of the hub was still adhered to the needle cannula.Microscopic examination revealed that the point of separation was smooth.It was also confirmed that the hub on the returned sample is the new hub design, which matches the bill of materials.Functional inspection was not able to be performed due to the damage to the needle hub.A device history record review was performed with no relevant findings.The report of a separated needle hub was confirmed by complaint investigation of the returned sample.Based on the sample provided, design likely caused or contributed to this event.A capa was opened to further investigation this issue.Corrective actions have not yet been implemented.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 3-LUMEN 7 FR X 20 CM
Common Device Name
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key14501638
Report Number3006425876-2022-00489
Device Sequence Number5124211
Product Code DQY
UDI-Device Identifier10801902157970
UDI-Public10801902157970
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K862056
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberIPN914770
Device Catalogue NumberCV-15703
Device Lot Number71F21F0035
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/25/2022
Supplement Report FDA Received Date06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-