U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIGFOOT BIOMEDICAL, INC. BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIGFOOT BIOMEDICAL, INC. BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM Back to Search Results
Model Number FG-300185
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypoglycemia (1912); Dizziness (2194); Cognitive Changes (2551); Balance Problems (4401)
Date of Event 04/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer narrative: no device issues were reported by the customer.Bigfoot conducted a thorough investigation and there was no indication that the product did not meet specification.Data logs from the patient's unity system were reviewed and showed that the system issued a low glucose alert per specification and the alert was acknowledged.After initial acknowledgement, the system issued alerts again when the customer's glucose reached the low and very low levels and these alerts resolved per specification.If bigfoot learns of any new information in relation to this case, another investigation will be performed, and a follow-up report will be submitted.All pertinent information available to bigfoot has been submitted.
 
Event or Problem Description
A customer experienced symptoms of severe hypoglycemia including lightheadedness, stumbling/falling, feeling lethargic, and having a hard time remembering.The customer required assistance from their step-son who provided some carbohydrates (hamburger steak, salad, and slaw) to treat the low glucose condition.There was no report of death or permanent impairment associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIGFOOT UNITY DIABETES MANAGEMENT SYSTEM
Common Device Name
DIABETES MANAGEMENT SYSTEM
Manufacturer (Section D)
BIGFOOT BIOMEDICAL, INC.
1820 mccarthy boulevard
milpitas CA 95035
Manufacturer (Section G)
BIGFOOT BIOMEDICAL, INC.
1820 mccarthy boulevard
milpitas CA 95035
Manufacturer Contact
kate lee
1820 mccarthy boulevard
milpitas, CA 95035
4087165600
MDR Report Key14504418
Report Number3016525500-2022-00005
Device Sequence Number14729014
Product Code QLG
UDI-Device Identifier00850003506227
UDI-Public00850003506227
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K202145
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2022
Device Model NumberFG-300185
Device Catalogue NumberFG-300185
Device Lot Number2590
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Initial Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexFemale
-
-