The reported field event of premature battery depletion, failure to interrogate, data problem and loss of pacing were not confirmed.The device was received in backup operation.The device image showed code corruption and this was the cause of the backup operation.Analysis performed after reloading product code indicated normal device characteristics, voltage is still above elective replacement indicator (eri) level.Device was able to interrogate, and no anomaly was noted on pacing.Code corruption could not be reproduced.The device longevity assessment was performed and the device was found to be in the normal range of operation with appropriate remaining longevity.
|