• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 32MM -3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 32MM -3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 71343203
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a thr on (b)(6) 2018, the patient experienced dislocation on the right side.A revision surgery was performed on (b)(6) 2018 to revise the oxinium fem hd 12/14 32mm -3.Outcome recovered and resolved.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation.The clinical/medical investigation concluded that, the finite data provided in the case report forms support the complaint; however, the limited information was insufficient to base an assessment of the root cause and patient impact beyond the reported events.Should clinical documentation become available in the future, a thorough medical assessment may be rendered at that time.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that patients should be warned that unassisted activity, particularly activities requiring excessive motion of the hip, may result in dislocation.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.The contribution of the device to the reported incident could be corroborated as it was required a revision surgery to treat this event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
Correction in g3 from initial report.It was stated that the date received by manufacturer was 09-may-2021, but there was a typo during web intaking it, should have been 09-may-2022.This information was clarified by the initial reporter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 32MM -3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14507145
Report Number1020279-2022-02596
Device Sequence Number12153789
Product Code LZO
UDI-Device Identifier03596010474216
UDI-Public03596010474216
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source Other,Foreign,Study,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71343203
Device Catalogue Number71343203
Device Lot Number17AM10130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/09/2021
Supplement Date Received by Manufacturer07/05/2022
07/05/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date07/05/2022
11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/13/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
R3 0 DEG XLPE ACET LNR 32MM X 52MM-17FM01823; R3 MULTI HOLE ACETABULAR SHELL 52MM-17KM03511; SL-PLUS INTEGR MIA STEM LAT. W. TI/HA 3-B1615754
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient SexFemale
Patient Weight44 KG
-
-