• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 PHASIX MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 PHASIX MESH; SURGICAL MESH Back to Search Results
Catalog Number 1191025
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Wound Dehiscence (1154); Impaired Healing (2378)
Date of Event 04/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, about 11 months post-implant of the phasix mesh, the patient was diagnosed with wound dehiscence and mesh exposure.As reported the ae has not recovered/not resolved.The clinician has assessed the patient¿s postoperative course as being ¿definitely¿ related to the study device and procedure, however, based on the information provided, no conclusion can be made.Wound dehiscence is a known inherent risk of surgery and is listed in the instructions-for-use supplied with the device as a possible complication.Review of manufacturing records confirms product was manufactured to specification.Not returned - remains implanted.
 
Event or Problem Description
As reported per prevent clinical trial dvl-he-018: (b)(6).The subject patient underwent a primary laparotomy procedure to perform a choleyeystocolonic fistula takedown, in which the bard/davol phasix mesh was placed in onlay fashion.The phasix mesh was not trimmed to size, with the appropriate overlap in all directions beyond the margins of the fascial incision.Mechanical fixation (bard/davol sorbafix) was used to fixate the mesh.The midline fascia was completely closed with non-bard/davol, 1pds slow absorbing monofilament sutures in a running stitch technique.Skin closure was achieved using long-term absorbable sutures.(b)(6) 2021, the subject patient was discharged from the hospital.(b)(6) 2022, the patient experienced a wound complication with exposure of the phasix mesh."wound in the distal area is approx.4cm in size in original incision, scabbed over, but patient states scab falls off and re-scabs.Mesh is slightly exposed.Wound in proximal area approx.3cm, no mesh." the reported adverse event has been assessed per clinician as definitely related to the study device and definitely related to the procedure.As reported, the outcome of the adverse event not recovered/not resolved.The severity is reported to be mild with no action taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHASIX MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14513674
Report Number1213643-2022-00345
Device Sequence Number17533788
Product Code OOD
UDI-Device Identifier00801741107580
UDI-Public(01)00801741107580
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K161424
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Study,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2022
Device Catalogue Number1191025
Device Lot NumberHUEV1630
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/02/2022
Initial Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
Patient Weight107 KG
-
-