BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE ID; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006L8406921
Medical Device Problem Codes Break (1069); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation results: the returned flexiva pulse id laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in one piece.The laser fiber body measured 277.5 cm from the tip of the sma connector to the distal end of the distal tip.There was no exposed glass tip present, the tip was manually clipped off.A functional evaluation revealed that the light from the sma connector was distorted, indicating a fracture within the connector.No other problems with the device were noted.The reported event of the flexiva pulse id laser fiber failed to emit energy was confirmed.It is likely that procedural handling of the fiber during set-up or use contributed to the fracture within the fiber connector and fiber body fracture.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that a flexiva pulse id laser fiber was used in a ureteroscopy procedure for a stone treatment in the ureter performed on (b)(6).During the procedure, the laser fiber did not work.The procedure was completed with another flexiva pulse id laser fiber.There were no patient complications reported as a result of this event.Note: this event has been deemed reportable based on the investigation finding of laser fiber broken within the connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA PULSE ID
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14514687
Report Number3005099803-2022-02925
Device Sequence Number6466287
Product Code GEX
UDI-Device Identifier08714729978732
UDI-Public08714729978732
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K210925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2025
Device Model NumberM006L8406921
Device Catalogue Number2130-02
Device Lot Number0028731378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/06/2022
Initial Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-