Model Number IOSM313
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 04/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint (b)(4).A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant was removed due to bone loss.
 
Additional Manufacturer Narrative
Multiple mdr reports are filed for this event.See 0001038806-2022-00780, 0001038806-2022-00780-1, 0001038806-2022-00778, and 0001038806-2022-00778-1.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhrs.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar events and 2 other relevant complaints were identified.As documented in the summary investigation, contributing factors for the reported event likely exist outside of zimmer biomet control, including those related to medical conditions / patient habits and surgical technique.Based on the summary investigation, no malfunction occurred upon investigation and no association between the dental implant and bone loss was found that can explain these events.Therefore, the reported event remains non-verifiable as it is a medical condition.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® CERTAIN® IMPLANT 3.25 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14515412
Report Number0001038806-2022-00781
Device Sequence Number5628892
Product Code DZE
UDI-Device Identifier00844868007043
UDI-Public(01)00844868007043(17)211216(10)2016110196
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/16/2021
Device Model NumberIOSM313
Device Catalogue NumberIOSM313
Device Lot Number2016110196
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-