MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00539330
Medical Device Problem Codes Break (1069); Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a flexima biliary stent was used during a biliary drainage procedure in the bile duct performed on (b)(6) 2022.During the preparation, when the device was unpacked, it was found that the delivery system was not connected to the stent.The procedure was successfully completed with another flexima biliary stent.There were no patient complications reported as a result of this event.Investigation results revealed that the guide catheter was detached/separated, therefore this event has been deemed an mdr reportable event.
 
Additional Manufacturer Narrative
(b)(4).The returned flexima biliary stent was analyzed, and a visual evaluation noted that the guide catheter was detached from the proximal section.The stent remained attached to the delivery system by the suture thread.Additionally, it was observed that the suture hole of the push catheter was torn and the guide catheter at the distal part was observed stretched.The push catheter was kinked at some sections.No other problem with the device was noted.The reported event was not confirmed.Based on the provided and collected information, this device met all manufacturing requirements and no abnormalities were reported during the manufacturing process; after analysis, it was determined that the delivery system was manipulated, the push catheter was kinked, the guide catheter was detached at the proximal part, the stent remained attached to the delivery system indicating that there was not a premature deployment, instead it seems that there was a failure deployment, the suture hole of the push catheter was torn and the guide catheter was stretched.It is possible that user manipulation during preparation and/or some technique performed with the delivery system caused the push catheter to get kinked.As a consequence of these kinks, the user may have experienced difficulties to remove the guide catheter and this resulted in detaching the proximal part of the guide catheter and stretching part of it.The tough movement experienced during these actions may have caused the tear in the push catheter suture hole.Based on the information available and the analysis performed, it was determined that the overall investigation conclusion code for the complaint will be documented as "adverse event related to procedure" since the adverse event occurred during the procedure and the device had no influence on event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14515605
Report Number3005099803-2022-02756
Device Sequence Number6466318
Product Code FGE
UDI-Device Identifier08714729162643
UDI-Public08714729162643
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K965147
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2024
Device Model NumberM00539330
Device Catalogue Number3933
Device Lot Number0027352943
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/03/2022
Initial Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
Patient Weight58 KG
-
-