It was reported that a minimum fill notification occurred after the user filled the cartridge with 100 units of insulin during the load sequence.A new cartridge was loaded to resolve the issue.Reportedly, customer then received cartridge change errors occurred with multiple cartridges during the load sequence and reported that a cartridge did not fit onto the pump.The customer did not have an alternate method of insulin therapy available but declined follow up from tandem technical support to confirm an alternative method of insulin therapy was obtained.Customer¿s blood glucose level was 296 mg/dl.
|