BOSTON SCIENTIFIC CORPORATION ADVANIX BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00534210
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix rx biliary stent was used during an endoscopic retrograde cholangiopancreatography procedure, on an unknown date.On (b)(6) 2022, a planned stent removal procedure was performed.When the physician attempted to remove the stent, a piece of the stent broke off.The procedure was completed and it was unknown if a new stent has been implanted.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that an advanix rx biliary stent was used during an endoscopic retrograde cholangiopancreatography procedure, on an unknown date.On (b)(6), 2022, a planned stent removal procedure was performed.When the physician attempted to remove the stent, a piece of the stent broke off.The procedure was completed and it was unknown if a new stent has been implanted.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block d4, h4: the complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Block h6: medical device problem code a0401 captures the reportable event of stent break.Block h10: the returned advanix rx biliary stent was analyzed, and a visual evaluation noted that, it was detached, kinked, and stretched at the detached section.The delivery system was not returned for analysis.No other issues with the device were noted.The reported event was confirmed.According to the product analysis, it is most likely that during the removal procedure the technique applied and/or some anatomical factors contributed on kinking the stent and it is possible that the kink caused resistance while removing it, as consequence the customer may have applied a different technique or tension to remove the stent and this could have caused the stretch and detachment of the stent.Therefore, adverse event related to procedure is selected as the most probable root cause for the complaint.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14515821
Report Number3005099803-2022-02822
Device Sequence Number15112088
Product Code FGE
UDI-Device Identifier08714729787303
UDI-Public08714729787303
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K101314
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00534210
Device Catalogue Number3421
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-