MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114362
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that after treatment with an ak 96 machine, an external fluid leak was observed.It was further reported that there was a blue bypass tube broken due to aging.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: h3, h6, h10.The device was not received for evaluation; however, the device was evaluated on site by a hospital technician.Upon inspection, a broken blue dialyzer tube was observed to be leaking.The reported condition was verified.The component was replaced, and the device was returned back to service with no issues noted.The cause of the condition was due to aging of the blue dialyzer tube.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14515898
MDR Text Key300354935
Report Number9616026-2022-00054
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114362
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-