MEDOS INTERNATIONAL SARL CH 5.5 TI CORT FIX 7X35MM; BONE-SCREW INTERNAL SPINAL FIXATION SYSTEM, NON-STERILE
|
Back to Search Results |
|
| Model Number |
186731735 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter occupation: reporter is a j&j sales representative.The lot number was unknown.Therefore, the expiration date and device manufacture date were unknown.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
|
| |
|
Event or Problem Description
|
|
Device report from depuy synthes reports an event in japan as follows: it was reported by the sales rep in japan that during a tlif (l4-5-s) surgery on (b)(6) 2022 for treating lumbar spinal canal stenosis with viper 2 system, an open extender was attached to the screw in question for reduction.According to the report, when they tried to deploy a rod, the l5 extender (unk) came off.It was reported that the screw in question was removed, and it was confirmed that the screwhead had been damaged.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.This complaint involves one (1) device.
|
| |
|
Event or Problem Description
|
|
This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D7: it is unknown if this single-use device was reprocessed and reused.
|
| |
|