• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FILTERWIRE EZ; TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION Back to Search Results
Model Number 7576
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that the wire broke.A 300cm filterwire ez was selected for use.During the preparation, it was noted that the wire broke when trying to expel the air from the basket.The procedure was completed with a different device.There were no complications reported.
 
Additional Manufacturer Narrative
E1.Initial reporter address 1: (b)(6) lab.Device evaluated by manufacturer: the device was returned for analysis.The wire returned inside the delivery sheath and the filter bag came back undeployed.It is possible to observe the spring tip is bent and stretched.Also the wire was exposed through the sheath slit.No detachment sections were noted.Under microscope, it was observed that the spring tip is bent and stretched.
 
Event or Problem Description
It was reported that the wire broke.A 300cm filterwire ez was selected for use.During the preparation, it was noted that the wire broke when trying to expel the air from the basket.The procedure was completed with a different device.There were no complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Common Device Name
TEMPORARY CORONARY SAPHENOUS VEIN BYPASS GRAFT FOR EMBOLIC PROTECTION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14516535
Report Number2134265-2022-05842
Device Sequence Number5502818
Product Code NFA
UDI-Device Identifier08714729470861
UDI-Public08714729470861
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K063313
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2023
Device Model Number7576
Device Catalogue Number7576
Device Lot Number0028507984
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/07/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date06/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-