Model Number OTP5000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheters are defective.Verbatim: here are the details around the otp5000 product i retrieved from both campuses: 8 each - john muir concord campus, lot# 0001441008.2 each - john muir concord campus, lot# 0001437013.Per sales rep not all products broke during use.Products were defective prior to use, however sales rep was unable to give specifics.Attached is the video sent to me from a physician who went to use the otp5000 thoracentesis catheter.As you can see he is able to disconnect the blue catheter from the hub quite easily.
 
Additional Manufacturer Narrative
(b)(4).Initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
 
Event or Problem Description
It was reported that catheters are defective.Here are the details around the otp5000 product i retrieved from both campuses: 8 each - john muir concord campus lot# 0001441008.2 each - john muir concord campus lot# 0001437013.Per sales rep not all products broke during use.Products were defective prior to use, however sales rep was unable to give specifics.Attached is the video sent to me from a physician who went to use the otp5000 thoracentesis catheter.As you can see he is able to disconnect the blue catheter from the hub quite easily.
 
Additional Manufacturer Narrative
Pr (b)(4) follow-up emdr for device evaluation: ten samples and one video were provided to our quality team for investigation.Upon visual inspection of these trays no issues were noted therefore, the reported failure mode was not confirmed.The video confirms the product drawing statement that the catheter must meet force at break requirements.A review of the internal manufacturing device records and raw material history files for the reported lot numbers 0001441008 & 0001437013 were performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Based on the available information we are not able to identify a root cause at this time.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and defect will continue to be monitored by our quality team for signs of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORA/PARA TRAY, 5FR- NON-VALVED
Common Device Name
THORACENTESIS TRAY
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key14516709
Report Number1625685-2022-00062
Device Sequence Number12394704
Product Code PXI
UDI-Device Identifier10885403284748
UDI-Public(01)10885403284748
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberOTP5000
Device Catalogue NumberOTP5000
Device Lot Number0001437013
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/16/2022
Supplement Date Received by Manufacturer06/14/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-