Visual, dimensional, material and functional analysis could not be performed as the device was not returned.However, an x-ray image was provided.It cannot be determined if a slip occurred from the provided x rays as the x rays are at different angles.There appears to be no discernable slip from the images provided.Every other level is fixated alternatively on the construct.The construct ends at the convex curve of the deformity.Device history records could not be reviewed as a valid lot number was not provided and could not be obtained.Complaint history records were reviewed for this catalog number, similar complaints were identified.Similar construct design x rays from the same surgeon for the same failure mode were provided and reviewed by a consultant spine hcp.The hcp believes construct design with difficult deformities was the most likely contributing factor to the event.From his analysis, the surgeon used a low density screw construct ending the construct on the convex curve of the deformity.The low screw density, short length of construct along with possible loss of correction may not have been enough to counter the force of the spine trying to revert to it's original position, causing a failure at the construct's weakest point, the distal most screw on the convex of the deformity curve, leading to the rod slipping from the screw.Why it cannot be confirmed if a slip occurred from the images provided.Additionally, these devices were implanted for 1.5 years.The ifu states "internal fixation devices are load sharing devices which maintain alignment until healing occurs.If healing is delayed or does not occur the implant could eventually break, bend or loosen.Loads produced by load bearing and activity levels will impact the longevity of the implant.".
|