• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TOXO IGM; ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TOXO IGM; ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII Back to Search Results
Model Number TOXO IGM
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation reviewed the customer's calibration results and determined the results were acceptable.The investigation reviewed the provided qc results, which were not from the date of the event, and the results were within acceptable ranges.The sample from the patient was requested for further investigation, but no samples were available.Based on the provided information, the investigation did not identify a product problem.The cause of the event could not be determined.
 
Event or Problem Description
The initial reporter received questionable elecsys toxo igm results for one patient tested on a cobas 6000 e 601 module serial number 32l6-09.On (b)(6) 2022, the patient's initial elecsys toxo igm result was reported outside the laboratory.On (b)(6) 2022, the patient went to a reference center for further testing.On (b)(6) 2022, the patient had a new sample collected and the customer performed testing with the e 601 module.On (b)(6) 2022, the patient's initial elecsys toxo igm result was positive.The specific value was requested but not provided.On (b)(6) 2022, the patient's toxo igm result at the reference center and by "cmia" methodology was negative.On (b)(6) 2022, the patient's elecsys toxo igm result was 5.7 coi positive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TOXO IGM
Common Device Name
ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14516872
Report Number1823260-2022-01500
Device Sequence Number5628928
Product Code LGD
UDI-Device Identifier04015630921485
UDI-Public04015630921485
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K162678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model NumberTOXO IGM
Device Catalogue Number04618858190
Device Lot Number59034001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Initial Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR
Patient SexFemale
-
-