MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LEVEEN SUPERSLIM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 36705
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle was bent.A 3.0/17/15 leveen superslim needle electrode was selected for use in a tumor ablation procedure.During the procedure, the needle became bent and the console frequently shutdown.The procedure was completed with another of the same device.No patient complications were reported, and the patient was in stable condition following the procedure.
 
Additional Manufacturer Narrative
Device eval by manufacturer: one rf probe electrode was returned for analysis.It was observed two tines were bent at the distal section.A functional evaluation extended and retracted the tines without resistance.
 
Event or Problem Description
It was reported that the needle was bent.A 3.0/17/15 leveen superslim needle electrode was selected for use in a tumor ablation procedure.During the procedure, the needle became bent and the console frequently shutdown.The procedure was completed with another of the same device.No patient complications were reported, and the patient was in stable condition following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEEN SUPERSLIM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14517208
Report Number2134265-2022-06030
Device Sequence Number12163760
Product Code GEI
UDI-Device Identifier08714729441717
UDI-Public08714729441717
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113092
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2024
Device Model Number36705
Device Catalogue Number36705
Device Lot Number0028432945
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer07/04/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
Patient Weight60 KG
-
-