| Catalog Number |
07.02030.004 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the screws on a crosslink are jammed.There were no patient impacts.
|
| |
|
Event or Problem Description
|
|
It was reported that the screws on a crosslink are jammed.There were no patient impacts.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed no signs of obvious damage.A functional inspection was performed on the transverse connector (pn 07.02030.004), attempting to turn the screws with a set screw starter (07.02119.001) found that the screw on one of the jaws is jammed, unable to turn at all.Root cause: root cause was unable to be determined.This event could possibly be attributed to cross-threading or applying an off-axis force.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Search Alerts/Recalls
|