• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS MINI STICK MAX; VESSEL DILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS MINI STICK MAX; VESSEL DILATOR Back to Search Results
Catalog Number 35506
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
A risk manager reported an issue with a nvi mini stick 6f x 20cm.018 ni/tu 21g chiba echo tip, via user facility medwatch.It was reported that the end user was attempting to push the wire through the access needle, and when doing so, a part of the "cope wire" broke off into two piece's inside of the patient's hepatic vein.It is unknown if the wire had been pulled back against the access needle when attempting to push it through.The two wire fragments were retained inside the patient as the provider was unable to remove them.The procedure was not completed due to this event.It was reported the patient was discharged from the hospital with hospice care, not related to the case.
 
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of part of the core wire broke off into two piece's inside of the patient's hepatic vein was not confirmed.No guidewire sample was returned for evaluation.A definitive root cause for this reported complaint description cannot be determined.The guidewire component is supplied to angiodynamics by the supplier/manufacturer heraeus medical.Scar004565 was sent to heraeus medical for awareness of this event and dhr review of supplier lot {9008425220}.As per scar004565 response, the review of the dhr demonstrated compliance and conformance to applicable procedures and specifications.General training review was performed in the sap system.As a result, it was confirmed that all personnel was properly trained in all manufacturing process steps and quality inspections as well.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use {dfu} is provided with this mini stick coaxial microintroducer.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events: guidewire shearing, fracture or embolization.Operational instructions: 1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI STICK MAX
Common Device Name
VESSEL DILATOR
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key14517851
Report Number1317056-2022-00082
Device Sequence Number6635659
Product Code DRE
UDI-Device IdentifierH965355061
UDI-PublicH965355061
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K132713
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number35506
Device Lot Number5724767
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date06/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient Weight61 KG
-
-