| Model Number |
N/A |
| Medical Device Problem Codes |
Break (1069); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refz1191 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported "perfectly in the center of the vessel.Advanced wire without resistance and advanced catheter threaded without resistance at the very end slight resistance.Went to safety the needle and resistance.Couldn¿t pull back.Hub of catheter broke from catheter.Had to use forceps to get needle and guide wire out.Catheter stuck in patient forearm.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged midline catheter was confirmed.The product returned for evaluation was one 22ga x 8cm powerglide pro midline catheter assembly.The catheter had been advanced and the safety mechanism was engaged over the needle tip.The catheter shaft exhibited a complete break dust distal of the molded joint.The distal catheter fragment was not returned for evaluation.The break site appeared irregular.Microscopic inspection of the break site revealed a partially glossy and partially dully granular fracture surface.The break exhibited a tapered profile.The break features were consistent with the catheter being damaged by contact with the introducer needle tip.Such damage can occur if the catheter is pulled back onto the needle and if the needle is re-inserted following catheter advancement.
|
| |
|
Event or Problem Description
|
|
It was reported "perfectly in the center of the vessel.Advanced wire without resistance and advanced catheter threaded without resistance at the very end slight resistance.Went to safety the needle and resistance.Couldn¿t pull back.Hub of catheter broke from catheter.Had to use forceps to get needle and guide wire out.Catheter stuck in patient forearm.".
|
| |
|