| Model Number |
442970 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while testing with bd max¿ ct/gc/tv 20 false negative results for tv target occurred over 6 month period.Pap smears obtained positive results for tv on the same samples.There is no report of patient impact.The following information was provided by the initial reporter: customer reported to bd-marketing 20 false negative specimens for tv target over a 6 month period.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that while testing with bd max¿ ct/gc/tv 20 false negative results for tv target occurred over 6 month period.Pap smears obtained positive results for tv on the same samples.There is no report of patient impact.The following information was provided by the initial reporter: customer reported to bd-marketing 20 false negative specimens for tv target over a 6 month period.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the complaint investigation for discrepant result when using the bd max ct/gc/tv (ref 442970) from unknown lot was performed by the verification of complaints history.Despite multiple attempts made to receive information from the customer, no data was provided for the investigation.The investigation was thus limited.Customer reported that, over a period of 6 months, 20 out of 300 samples gave a negative result with the bd max¿ ct/gc/tv assay whereas the pap smear test showed a positive result for the tv target.Based on the available information, and without run data, bd is unable to identify a cause for the customer discrepant results.Nonetheless, no reagent issue is suspected.There is no indication of a reagent issue based on the analysis of the complaints received for discrepant results on bd max ct/gc/tv products.The root cause was not identified.Bd cannot confirm the complaint based on the investigation that was performed.H3 other text : see h10.
|
| |
|