As reported, the (b)(6) male patient had a revision due to stem subsidence.Patient was revised with a zb stem, cup was retained and a new liner was implanted.There was no breakage of a device.There was no surgical delay/prolongation.Patient was last known to be in stable condition following the event.Photos and x-rays received.Product will not be returned.
|
|
After further review of additional information received the following sections d4, d6a, d10, g3, g6, h2, h3, h4 and h6 have been updated accordingly.Section d10: concomitant products: nv gxl lnr, +5lat, 36mm g3-56/58mm cups (cat# 136-36-53 / serial# (b)(6)).(h3) the revision reported may have been the result of either under sizing the femoral stem for the patient¿s anatomy/femoral bone and/or insufficient bony support for the implant, possibly due to a patient-related condition, which led to femoral stem subsidence.However, this cannot be confirmed as immediate post-operative x-rays of the index surgery were unable to be obtained.Therefore, a change in the position of the femoral stem from initial implantation cannot be confirmed.The most probable root cause associated with the reported event of "femoral stem subsidence¿ is associated with a femoral stem that subsides into the prepared femoral canal further than expected.This can occur during implantation or post-operatively.
|