TENEX HEALTH, INC. TXB TISSUE REMOVAL SYSTEM - MICROTIP; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Model Number 554-3003-001
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 04/22/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During a procedure using the tx microtip, smoke was visible from the needle insertion point.The device was removed from the patient and a skin burn was noted.Burn and pain treatments were administered to treat the patient.
 
Additional Manufacturer Narrative
Follow-up report #01.Sections b4, d9, g6, h2, h3, h6 and h10 have been updated with new information.The device was returned and evaluated.Due to the state in which the device was received, functional testing could not be performed.Disassembly of the device revealed a local point of damage on the outside of the microtip needle near the open end of the outlying case nose tube.It appears the needle had been pushed against the case nose, leading to friction and local heat during operation.It is unclear what specifically led to the side loading / angling of the needle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TXB TISSUE REMOVAL SYSTEM - MICROTIP
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630 8123
Manufacturer (Section G)
TENEX HEALTH, INC.
26902 vista terrace
lake forest CA 92630
Manufacturer Contact
david vancelette
26902 vista terrace
lake forest, CA 92630
9492380822
MDR Report Key14519427
Report Number1000135560-2022-00011
Device Sequence Number15059971
Product Code LFL
UDI-Device Identifier00857156006093
UDI-Public(01)00857156006093(10)2232101(17)230811
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K181367
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/10/2023
Device Model Number554-3003-001
Device Catalogue Number554-3003-001
Device Lot Number2222102
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age65 YR
Patient SexMale
Patient Weight100 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-