GYRUS ACMI, INC CYSTOURETHROSCOPE STANDARD OBTURATOR, 23FR; RIGID CYSTOURETHROSCOPE Back to Search Results
Model Number E123-O
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not yet been received for evaluation.The cause of the issue is unknown at this time.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
Customer reported with an issue of " the tip is broken off".The issue occurred during reprocessing.There was no patient harm , no user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The initial reporters occupation was inadvertently entered into the initial medwatch report incorrectly.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Upon completion of the evaluation, there were no problems found with the distal tip.However, the obturator knob at the proximal end was found damaged, and the red dot on the release button was missing.It¿s likely the damaged knob was due to improper handling during device reprocessing.The root cause of the damaged knob was unable to be identified.The device was manufactured in june 2016.However, the specific date is unknown.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOURETHROSCOPE STANDARD OBTURATOR, 23FR
Common Device Name
RIGID CYSTOURETHROSCOPE
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14519880
Report Number8010047-2022-09004
Device Sequence Number4793728
Product Code FEC
UDI-Device Identifier00821925002371
UDI-Public00821925002371
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE123-O
Device Lot NumberFW
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-