• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE® CERTAIN® IMPLANT 3.25 X 13MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE® CERTAIN® IMPLANT 3.25 X 13MM; DENTAL IMPLANT Back to Search Results
Model Number IOSM313
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
It was reported that the implants was removed due to medical bone loss.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).A device history record review was performed and no related deviations or nonconformances were noted.Also, a complaint history search was performed using our complaint handling system and there were two related complaints for this product lot.Complainant reported that the implant was removed due to bone loss at the implant site.The reported complaint could not be verified.A definitive root cause for this complaint could not be determined.Bone loss: a summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.The following sections have been updated: d9: device available for evaluation change ¿no' to 'yes'.H3: device evaluated by manufacturer: change ¿no' to 'yes'.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® CERTAIN® IMPLANT 3.25 X 13MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key14520092
Report Number0001038806-2022-00795
Device Sequence Number15069197
Product Code DZE
UDI-Device Identifier00844868007043
UDI-Public(01)00844868007043(17)211216(10)2016110196
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/16/2021
Device Model NumberIOSM313
Device Catalogue NumberIOSM313
Device Lot Number2016110196
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date05/26/2022
Supplement Report FDA Received Date10/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-